Black Body Source - Low & Medium Temperature Combination
Model : CN-DT
Portable Dual Black Body source in a Single Unit - combination of Low & Medium Black Body sources to cover temperature ranges from Ambient to 300°C (Low) & 250 to 600°C (Medium) with an high accuracy.
Easy to use, it is fitted with controller that displays simultaneously the setpoint and the measured temperature.
- Features
- Specifications
- Datasheet
- Standard delivery
- Enables simultaneous measurement of 2 Pyrometers
- Quick Measurement in 2 Points over the range
- Compact, Light Weight, Bench-top & Portable
- Dual Ranges : 50 to 300°C & 250 to 600°C
- Excellent Accuracy & Stability
- High Emissivity, wide Aperture
- Auto-tune PID Controller
- Low Power Consumption
- Cost effective - Best Value for Money
- RS-232 Computer Interface - Optional
- Calibration Management Software - Optiona
| Black Body System Temperature Calibrator | ||
|
CN-DT |
Chamber 1 |
Chamber 2 |
| Range | 50°C to 300°C | 250°C to 600°C |
| Accuracy |
< 1°C | < 2°C |
| Stability | ±0.5°C | |
| Resolution | 0,1°C | |
| Temperature control |
Auto-tune PID Controller | |
| Heating time (to max.) |
45 minutes | |
| Stabilization time | 30 minutes | |
| Temperature Readout | °C / °F selectable | |
| Emissivity | 0.98 | |
| Aperture size |
Ø 25 mm | |
| Power Supply |
230V AC / 50 Hz / 3000 W | |
| Dimension | 450 X 410 X 190 mm (W x D x H) | |
| Weight | 14 Kg | |
| Housing / Mounting |
Bench-top / Table-top (Portable) | |
- Basic Instrument
- Mains Cable
- Spare Fuses
- Traceable Calibration Certificate
- Instruction Manual
- RS-232 Computer Interface
Â
OPTIONAL
- "Caltemp" Calibration Software
- Contact Thermocouple (to measure cavity temperature)
- Thermocouple holding stand
- Calibration Certificate from NABL Accredited - as per ISO/IEC 17025:2005 Laboratory